Versati™ Micro Centrifuge

Versati™ Micro Centrifuge

View Other Laboratory Centrifuge

Distinct Control Panel and Intelligent Versati™ Microprocessor Control System

  • Big rotary knob and buttons, convenient for rotation and operation.
  • Large LCD screen with clear indication of running condition including rotor, speed and time.
  • Operation panel is water-proof, can well protect the inner electrical components.
  • Microprocessor with speed self-tested system.

Genuine-Protec™ Safety Lid Lock

  • Motor driven lid lock, ensures safety and reliability.
  • The lid can only be opened when rotor stops, and speed indicator is 0.
  • Smooth, automatic, and electromagnetic lid allows one-hand-operation

V-balance™ Weight Imbalance Protection

  • Equipped with active weight imbalance-identification system. In case the rotor is not equally loaded, the drive will turn off during acceleration. The rotor decelerates to stop.

Diverse Choices of Rotors

  • Made of aluminum alloy.
  • Powerful motor provides high RCF (Relative Centrifugal Force) up to 21380 x g for MCV/MCR.
  • As many as 7 rotors with wide range capacity for MCV and 6 rotors for MCR.

Integrated Design of Interior Centrifuge Chamber

  • The side plate and baseboard of centrifuge chamber are made by one-time punch forming of a whole piece of steel plate.
  • No dead angle within the chamber, easy to clean and can reduce pollution at the greatest possibility.

Wide Range of Temperature

  • Not only have a cooling function, but also a heating function. The temperature range is from -20OC to +40OC.

Maintenance-free, Brushless Motor

  • Equipped with German-originated motor, requiring neither bearing replacement nor change oil.
  • No carbon brush, low interference. Electric- spark-free that ensures no interference to radio equipment.
  • Low noise and smooth operation with superior mechanism design.
  • Prolongs the lifespan and save maintenance cost.

Environment-Friendly Refrigeration Technology

  • Designed with CFC-free refrigeration system for refrigerated models (MCR/TCR), cause no harm to the atmosphere.

Emergency Switch

  • In case of power failure or malfunction, the lid of the centrifuge can be opened manually via the emergency switch in order to protect your samples.

Polished and Practical Appearance

  • External surface are powder coated Esco ISOCIDE™ to eliminate 99.9% of surface bacteria within 24 hours of exposure. This technology ensures relatively aseptic condition.
  • Ensures healthier, safer and cleaner laboratory environment.

MODEL

VersatiTM MCV

VersatiTM MCR

Width

280 mm  (11”)

280 mm (11”)

Depth

389 mm (15”)

563 mm (22”)

Height

304 mm (12”)

294 mm (11.5”)

Weight Without Rotor

17 kg (37 lbs)

35 kg (77 lbs)

Max. Speed

(Revolution per Minute)

15000 rpm

Max. Volume

44 x 1.5/2.0 ml

Max. RCF

21380 x g

Rotor Options

7

6

Temperature Control Range

Ambient

-20°C to 40°C

Time Set Range

99 hrs 59 mins / 59 min 50 s in 1min / 10s increments / continuous

Allowable Kinetic Energy

7060 Nm

Main Power Connection AC

115V / 230 V; / 50-60 Hz

Voltage Fluctuation

± 10 %

Current Consumption

1.3 A/2.3 A

2.4 A/5.1 A

Power Consumption

270 W/280 W

500 W/560 W

Centrifuges are important laboratory equipment in most laboratories. To meet all needed applications, accessories enable to maximize the use of the centrifuge. To increase the versatility and performance, rotors and adapters are made available.

Rotors

The type of rotor commonly used in Micro Centrifuge is a fixed-angle rotor. This rotor can hold the container at a constant angle relative to the central axis. Choice of the rotor and lid will depend on the laboratory application.

 

Model No.

Max. Capacity

Angle

Max. RPM

Max. RCF

Remarks

VST-ROT-001

24 x 1.5 / 2.0 ml

45o

15000

21380 x g

 

VST-ROT-002

24 x 1.5 / 2.0 ml

45o

15000

21380 x g

Aerosol-tight

VST-ROT-003

30 x 1.5 / 2.0 ml

45o

13500

19151 x g

 

VST-ROT-004

44 x 1.5 / 2.0 ml

32oC/55oC

13500

17113 x g (outer slots) / 17317 x g (inner slots)

 

VST-ROT-006

12 x 5 ml

45o

14500

19978 x g

 

VST-ROT-007

4 x 8 (0.2 ml) PCR Strips

45o

15000

 15343 x g

For PCR Strips

VST-ROT-008

24 capillaries

90o

12000

14970 x g

Hematocrit Rotor for MCV only

Adapters

Adapters are designed to fit the vessels or tubes in the rotor bore and to avoid any spill due to shaking of unfit tubes.

 

Model No.

No. of Tubes and Adapter Capacity

Diameter (mm)

Length (mm)

VST -  ADP - 001

1 x 0.5 ml

8

45

VST – ADP - 002

1 x 0.4/0.2 ml PCR

6

45

VST – ADP - 003

1 x 1.5ml - 2.0 ml

11

42

VST – ADP - 004

1 x 1.0 ml Flat cryo tubes

12.6

41

VST – ADP - 005

1 x 1.8ml Dome cryo tubes

12.6

48